A fast, simple calibration method for organic carbon isotope analysis using continuous-flow elemental analyzer interfaced with an isotope ratio mass spectrometer.
Authors:Yeon-Sik Bong  Kwang-Sik Lee
Affiliation:Division of Isotope Geoscience, Korea Basic Science Institute, 52 Eoeun-dong, Yusung-gu, Daejeon 305-333, Korea.
Abstract:Continuous flow analysis using an elemental analyzer interfaced with an isotope ratio mass spectrometer (EA-IRMS) is faster and requires much less material than conventional analytical methods. Although using an EA-IRMS is simple and fast, accurate calibration strongly depends on matching sample and reference peak heights by adjusting the sample weight. This paper describes a new modification for calibration using only the major ion beam intensity (nA) without weighing each sample or increasing the number of reference materials.