Low solvent consumption gel permeation chromatography method
Authors:Adrian Boborodea  Stephen O'Donohue
Affiliation:1. Certech ASBL, Seneffe, Belgiumadrian.boborodea@certech.be;3. Agilent Technologies LDA UK Ltd., Craven Arms, Shropshire, UK
Abstract:This study presents the possibility of reducing the volume of solvent necessary for gel permeation chromatography analysis by up to 20 times when using two 2.1-mm internal diameter (ID) columns instead of the traditional three 7.5-mm ID columns. The low solvent consumption method was optimized by evaluating the number of theoretical plates at different flow rates. By injecting a broad polystyrene standard 16 times, good reproducibility and accuracy of weight average molecular weights (Mw) and peak surface areas were obtained for the optimized method.
Keywords:GPC  polystyrene  solvent consumption  tetrahydrofuran