Undistorted Cyclic Voltammograms at Scan Rates up to 2.5 MVs~(-1) through Positive Feedback Compensation of Ohmic Drop
作者姓名:郭智勇  林祥钦  邓兆祥
作者单位:Department of Chemistry,University of Science and Technology of China,Hefei,Anhui 230026,China,Department of Chemistry,University of Science and Technology of China,Hefei,Anhui 230026,China,Department of Chemistry,University of Science and Technology of China,Hefei,Anhui 230026,China
基金项目:the National Natural Science Foundation of China (No. 20173054).
摘    要:Introduction Though being the simplest voltammetric technique, cyclic voltammetry provides lots of electrochemical information especially in the study of the thermody-namic and kinetic processes and the detection of elec-trogenerated transient intermediates.1 Clearly, the more rapid the scan rate is employed, the faster the reactions can be studied. Indeed, a timescale RT/Fν~25 ns of the Faradaic response will be reached corresponding to a scan rateνof 1 MVs-1, which means that the method…


Undistorted Cyclic Voltammograms at Scan Rates up to 2.5 MVs~(-1) through Positive Feedback Compensation of Ohmic Drop
GUO,Zhi-Yong LIN,Xiang-Qin DENG,Zhao-Xiang.Undistorted Cyclic Voltammograms at Scan Rates up to 2.5 MVs~(-1) through Positive Feedback Compensation of Ohmic Drop[J].Chinese Journal of Chemistry,2004,22(9):913-919.
Authors:GUO  Zhi-Yong LIN  Xiang-Qin DENG  Zhao-Xiang
Institution:GUO,Zhi-Yong LIN,Xiang-Qin DENG,Zhao-XiangDepartment of Chemistry,University of Science and Technology of China,Hefei,Anhui 230026,China
Abstract:A circuit based on the current feedback operational amplifier was constructed to accomplish on-line ohmic drop compensation in ultrafast cyclic voltammetry. Firstly, its characteristics were confirmed experimentally on dummy cells. Then the reduction of anthracene in acetonitrile, a classical test example with very fast electron-transfer kinet-ics, was examined to prove them too. The results showed that this circuit could afford excellent ohmic drop com-pensation so that the undistorted voltammograms up to 2.2 MVs-1 scan rate can be recorded, and 2.5 MVs-1 if 5% error is tolerated.
Keywords:ultrafast cyclic voltammetry  megavolt per second scan rate  ultramicroelectrode  instrument  ohmic drop compensation