Abstract:Capillary Electrophoresis-Sodium Dodecyl Sulfate (CE-SDS) method with UV detection was developed and satisfactorily used for determination of purity and manufacturing consistency of a monoclonal antibody (MAb) at Amgen Inc. (Seattle, WA). When this method was applied to some other MAbs, several problems with method robustness became apparent. These issues resulted in abnormal Electropherogram (e-gram) profiles potentially linked to various parameters specific molecules analyzed, sample formulation buffer composition, CE-SDS gel matrix type, and operators. A multi-users interest group (called CE Users Forum) was formed to systematically investigate and understand these issues. The CE Users Forum first identified the issues which needed resolution, defined group experiments to better understand the problem and to test potential solutions, and together defined a generic (platform) CE-SDS method for MAbs. Two CE instruments, Agilent HP3DCE and Beckman PA 800, two CE-SDS gel matrices, BioRad and Beckman gels, as well as different types of MAbs in various buffers were used in this investigation. We present here a platform CE-SDS method for purity determination of MAbs. Method optimization and trouble-shooting procedures by the CE Users Forum played a key role in delivering a robust analytical method for characterization of antibodies by improving instrumental and experimental parameters such as instrument variability, instrument operating parameters, operator training, and reagent stability. The optimized CE-SDS method is used during process development and has been transferred to the quality control (QC) lab as a purity assay for lot release testing of therapeutic antibodies. Any trained analyst can successfully perform this method. A group such as the CE Users Forum is a good way to integrate best practices and solve technical issues in a cooperative environment.Presented at: CE in the Biotechnology & Pharmaceutical Industries: 7th Symposium on the Practical Applications for the Analysis of Proteins, Nucleotides and Small Molecules, Montreal, Canada, August 12–16, 2005
Keywords:Capillary electrophoresis-sodium dodecyl sulfate  Capillary gel electrophoresis  Monoclonal antibody separation  Troubleshooting  Optimization
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号