Analytical Strategy for the Determination of Selenium in Biological Materials by Inductively Coupled Plasma-Mass Spectrometry with a Dynamic Reaction Cell
Authors:Marcin Wieczorek  Anna Tobiasz  Danuta Dudek-Adamska  Stanis?aw Walas  Pawe? Ko?cielniak
Institution:1. Department of Analytical Chemistry, Faculty of Chemistry, Jagiellonian University in Kraków, Kraków, Polandmarcin.wieczorek@uj.edu.pl;3. Department of Analytical Chemistry, Faculty of Chemistry, Jagiellonian University in Kraków, Kraków, Poland
Abstract:Selenium was determined in serum, hair, and tobacco by inductively coupled plasma-mass spectrometry using 77Se, 78Se, and 82Se. The set of standards method (SSM) and the standard addition method (SAM) were applied to calibration with and without the use of internal standards (72Ge and 103Rh). In addition, the usefulness of the dynamic reaction cell (DRC) with methane as the reaction gas was characterized. The results obtained in different conditions were evaluated in terms of precision and accuracy. It was demonstrated theoretically and experimentally that an internal standard is a potential source of systematic errors as it can be influenced multiplicatively and additively by its own interferents (independently of selenium). Furthermore, it was shown that—against common opinion—an internal standard can fail in elimination of chemical interference effects influencing selenium and in increasing of precision of selenium determinations. The DRC was shown to be effective in the elimination of additive effects, although the results obtained by both SSM and SAM with DRC were systematically positively erroneous. Finally, selenium was determined accurately in each examined sample when SAM was applied to calibration, and signals were measured either for 82Se without the use of the DRC, or for 77Se or 78Se with the use of the DRC. In addition, it has also been shown that samples should be diluted prior to analysis to the greatest acceptable extent.
Keywords:Dynamic reaction cell  inductively coupled plasma-mass spectrometry  internal standard  set of standards method  standard addition method
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号