On-line preconcentration and determination of mercury in biological samples by flow injection vapour generation inductively coupled plasma atomic emission spectrometry |
| |
Authors: | Pedro Ca?ada Rudner José Manuel Cano Pavón Amparo García Torres Fuensanta Sánchez Rojas |
| |
Affiliation: | (1) Department of Analytical Chemistry, Faculty of Sciences, University of Málaga, E-29071 Málaga, Spain |
| |
Abstract: | An FI-ICP-AES method for the determination of trace levels of mercury in biological samples has been described, which is based on the extraction of the mercury complex with 1,5-bis (di-2-pyridyl)methylene thiocarbonohydrazide (DPTH) on-line into isobuthyl-methyl ketone (IBMK). The organic phase (containing the complex) has been mixed on-line with SnCl2 in N,N-dimethylformamide. Thus, mercury vapour can be generated directly from the organic phase and separated in a gas-liquid separation device. The detection limit for mercury is 4 ng/ml and the calibration curve is linear at least from 10 to 2500 ng/ml. The relative standard deviation for 10 replicate measurements is ±1% for 100 ng/ml of Hg(II). Results from the analysis of some certified biological reference materials are given. |
| |
Keywords: | |
本文献已被 SpringerLink 等数据库收录! |
|