首页 | 本学科首页   官方微博 | 高级检索  
     


Investigation of different types of filters for atmospheric trace elements analysis by three analytical techniques
Authors:A. E. Ali  J. Bacsó
Affiliation:(1) Institute of Nuclear Research, P.O. Box 51, H-4001 Debrecen, (Hungary);(2) Present address: Nuclear Physics Department, College of Science, Minia University, Egypt
Abstract:Different atmospheric aerosol samples were collected on three types of filters. Disks of both loaded and clean areas of each kind of filter were investigated by XRF, PIXE and Scanning Electron Microscope (SEM) methods. The blank concentration values of the elements Al, Si, P, S, Cl, K, Ca, Sc, Ti, V, Cr, Mn, Fe, Co, Ni, Cu, Zn, As, Br and Pb in the three types of fiters are discussed. It is found that for trace elemental analysis, the Nuclepore membrane filters are the most suitable for sampling. These have much lower blank element concentration values than the glass fibres and ash free filters. It was found also that the PIXE method is a more reliable analytical technique for atmospheric aerosol particles than the other methods.
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏