Evaluation of ODS-AQ stationary phase for use in capillary electrochromatography |
| |
Authors: | Djordjevic N M Fitzpatrick F Houdiere F |
| |
Institution: | Novartis Pharma, Core Technology Area, Basel, Switzerland. nebojsa.djordjevic@pharma.novartis.com |
| |
Abstract: | The aim of this study was to evaluate the applicability of ODS-AQ packing material as a stationary phase in capillary electrochromatography (CEC). The electroosmotic flow created on an ODS-AQ stationary phase was measured at different mobile phase compositions and at different column temperatures. It was observed that the electroosmotic flow generated in the column increased by 50% when the temperature of the system was raised from 20 degrees C to 60 degrees C, while all other conditions were kept constant. The electroosmotic flow produced by the ODS-AQ stationary phase was found to be comparable to the flow generated in a column packed with Nucleosil bare-silica material. In addition, a set of polar compounds (D-lysergic acid diethylamide derivatives) was utilized to determine the influence of temperature and mobile phase composition on their chromatographic behavior on an ODS-AQ stationary phase in a CEC mode. A linear relationship between the solute retention factor and column temperatures was seen over the temperature range studied (20 degrees C to 60 degrees C). A quadratic function was used to describe the changes in the solute retention factors with variation of acetonitrile concentration in the mobile phase. |
| |
Keywords: | |
本文献已被 PubMed 等数据库收录! |
|