首页 | 本学科首页   官方微博 | 高级检索  
     


The Development of Leukocyte Counter Using Fluorescence Imaging Analysis
Authors:Taisuke?Hirono  mailto:t-hirono@kowa.co.jp"   title="  t-hirono@kowa.co.jp"   itemprop="  email"   data-track="  click"   data-track-action="  Email author"   data-track-label="  "  >Email author,Katsumi?Yabusaki,Yukio?Yamada
Affiliation:(1) Research Laboratories, Electronics and Optics Res. Labs., Kowa Company, Ltd, 3-3-1, Chofugaoka, Chofu Tokyo, 182-0021, Japan;(2) Kowa Research Institute, Kowa Company, Ltd, Japan;(3) Dept. of Mech. Eng. and Intel. Sys, University of Electro-Communications, Japan
Abstract:
Image cytometry centrifugation (ICC) method has been developed and applied to count residual leukocytes in transfusion blood products for quality assurance testing. The standard leukocyte fluorescence image profile was established by averaging the actual fluorescence images of leukocytes. It was used for development of an optimum image analysis algorithm to avoid misinterpretation of the dust contaminated from the environment. The centrifugal concentration process of the ICC method was investigated by a computer simulation, and it was found that it caused leukocyte overlapping and limited the enumeration range. By introducing a double thresholding algorithm to the fluorescence images analysis and correcting the number of the overlapped leukocytes, we succeeded in improving the accuracy and the measurement range up to 25 [WBC/μl] of the ICC method.
Keywords:computer simulation  cytometry  fluorescence image  image analysis  leukocyte counter  transfusion blood product
本文献已被 SpringerLink 等数据库收录!