首页 | 本学科首页   官方微博 | 高级检索  
     


On-line purge and trap gas chromatography for monitoring of trihalomethanes in drinking water distribution systems
Authors:Brown Michael A  Miller Sarah  Emmert Gary L
Affiliation:The University of Memphis, Department of Chemistry, Memphis, TN 38152, USA
Abstract:
A method using an automated on-line purge and trap gas chromatograph with a dry electrolytic conductivity detector (DELCD) has been developed for monitoring four regulated trihalomethanes in drinking water distribution systems. This analyzer samples trihalomethanes from drinking water by pervaporation through a silicone capillary membrane contained within a gas extraction cell (GEC) followed by preconcentration using an adsorbent trap. Trihalomethanes are subsequently desorbed from the trap onto a capillary column, separated and detected. The analyzer operates in real-time, samples directly from the drinking water distribution system and is fully automated. The optimization, operation, and evaluation of the analyzer and method are discussed. Method detection limits (MDL) are less than 1.0 μg L−1 with acceptable estimates for accuracy, and precision. The results from two on-line monitoring studies in chlorinated and chloraminated distribution systems are presented. The performance of the method is compared directly to United Stated Environmental Protection Agency Method 502.2 and shows a very slight, but acceptable bias.
Keywords:Trihalomethanes   Drinking water   On-line monitoring   Gas chromatography   Disinfection by-products
本文献已被 ScienceDirect PubMed 等数据库收录!