首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Development of miniaturized pH biosensors based on electrosynthesized polymer films
Authors:Segut Olivier  Lakard Boris  Herlem Guillaume  Rauch Jean-Yves  Jeannot Jean-Claude  Robert Laurent  Fahys Bernard
Institution:a Institut UTINAM, CNRS-UMR 6213, Université de Franche-Comté, 16 route de Gray, Besançon Cedex 25030, France
b FEMTO-ST, CNRS-UMR 6174, Université de Franche-Comté, 32 avenue de l’Observatoire, Besançon Cedex 25044, France
Abstract:A new type of pH biosensor was developed for biological applications. This biosensor was fabricated using silicon microsystem technology and consists in two platinum microelectrodes. The first microelectrode was coated by an electrosynthesized polymer and acted as the pH sensitive electrode when the second one was coated by a silver layer and was used as the reference electrode. Then, this potentiometric pH miniaturized biosensor based on electrosynthesized polypyrrole or electrosynthesized linear polyethylenimine films was tested. The potentiometric responses appeared reversible and linear to pH changes in the range from pH 4 to 9. More, the responses were fast (less than 1 min for all sensors), they were stable in time since PPy/PEI films were stable during more than 30 days, and no interference was observed. The influence of the polymer thickness was also studied.
Keywords:Electrochemistry  pH biosensor  Microsystem technology  Polypyrrole  Linear polyethylenimine
本文献已被 ScienceDirect PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号