首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Sensitive determination of nucleotides and polynucleotides based on the fluorescence quenching of the Tb3+-Tiron complex
Authors:Dong-yuan Wang  Y Zhao  Jin-gou Xu and Xiang-qun Guo
Institution:(1) Department of Chemistry, The Research Laboratory of SEDC of Analytical Science for Material and Life Chemistry, Xiamen University, Xiamen 361005, China, CN
Abstract: A sensitive method using fluorescence quenching for the determination of nucleotides (ATP, ADP, AMP, CTP, UTP) and polynucleotidespoly(A), poly(I), poly(U)] is proposed. It is based on the ability of nucleotides and polynucleotides to inhibit the formation of a strongly fluorescent complex of Tb3+ ion with Tiron. The possibilities of spectrofluorimetric measurements of these systems were studied under optimal conditions (pH 6.9 in hexamethylene tetramine-HCl buffer, 1.2×10-6 mol/L of Tb3+, 4.0×10-6 mol/L of Tiron, λex=317 nm, λem=546 nm). The results showed that the Tb3+-Tiron complex could be used as a fluorescence test for the phosphate moieties of nucleotides and polynucleotides. The detection limits are 0.3, 1.2, 3.7, 0.2, 0.3, 1.1, 0.6 and 0.9 ng/mL for ATP, ADP, AMP, CTP, UTP, poly(A), poly(I), and poly(U), respectively. The relative standard deviations (6 replicates) are within 4.0% in the middle of the linear range. The fluorescence quenching mechanism of these systems is also discussed. Received: 16 July 1996 / Revised: 13 November 1996 / Accepted: 13 November 1996
Keywords:
本文献已被 SpringerLink 等数据库收录!