首页 | 本学科首页   官方微博 | 高级检索  
     


Pure Organic Phase Phenol Biosensor Based on Tyrosinase Entrapped in a Vapor Deposited Titania Sol‐Gel Membrane
Abstract:A novel amperometric biosensor was constructed for the determination of phenols in pure organic phase. This biosensor was fabricated by immobilizing tyrosinase in a titania sol‐gel membrane which was obtained with a vapor deposition method. This method was facile and avoided the calcination step needed in conventional titania sol‐gel process. The titania sol‐gel membrane could effectively retain the essential water layer around the enzyme molecule needed for maintaining its activity in organic phase. The experimental parameters such as solvent and operating potential were optimized. At ?100 mV this biosensor showed a good amperometric response to phenols in pure chloroform without any mediator and rehydration of the enzyme. For catechol determination the sensor exhibited a fast response of less than 5 seconds. The sensitivity of different phenols was as follows: catechol > phenol > p‐cresol. Additionally, the apparent Michaelis‐Menten constants of the encapsulated tyrosinase to catechol, phenol and p‐cresol were found to be 0.15±0.003, 0.17±0.008 and 0.21±0.004 mM, respectively. The biosensor had also good reproducibility and stability. This work provided a promising platform for the construction of pure organic phase biosensors and the determination of substrates with poor water solubility.
Keywords:Pure organic phase biosensor  Titania sol‐gel  Tyrosinase  Phenols  Vapor deposition