首页 | 本学科首页   官方微博 | 高级检索  
     检索      

ICP发射光谱测定保偏光纤应力棒中的二氧化二硼
引用本文:张淑珍,王春梅,吴琼.ICP发射光谱测定保偏光纤应力棒中的二氧化二硼[J].光谱学与光谱分析,1995(4).
作者姓名:张淑珍  王春梅  吴琼
作者单位:电子部46所
摘    要:本文提出了用ICP发射光谱仪测定保偏光纤应力棒中的二氧化二硼的方法。将应力棒粉碎后,使用HF在压力溶样器中消解,最后溶液的介质为0.5%氢氟酸,并采用耐腐蚀的样品引进系统进行ICP光谱测定。分析程序简单,准确度高,该方法回收率为99~101%,相对标准偏差为1.04%

关 键 词:ICP-AES,三氧化二硼,应力棒,加压分解

DETERMINATION OF B_2O_3 IN POLARIZATION-MAINTAINING OPTICAL FIBER STRESS APPLYING PART BY ICP-AES
ZHANG Shuzhen, WANG Chunmei and WU Qiong.DETERMINATION OF B_2O_3 IN POLARIZATION-MAINTAINING OPTICAL FIBER STRESS APPLYING PART BY ICP-AES[J].Spectroscopy and Spectral Analysis,1995(4).
Authors:ZHANG Shuzhen  WANG Chunmei and WU Qiong
Abstract:A method for the determination of B2O3 in polarization-maintaining optical fiber stress applying part by ICP-AES was studied. First we crushed the stress applying part.Then we dissolved it in pressure vessel with HF, and the solution to be analysed contained 0. 5% HF. Finally we performed the analysis with the Corrosion Resistant Sample Introduction System. The analysis procedure is very simple. The relative standard deviation is 1. 04%, and the recovery is 99-101 %.
Keywords:ICP-AES  Stress applying part  Pressure vessel  Boron oxide