首页 | 本学科首页   官方微博 | 高级检索  
     


Determination of carbaryl by flow injection with luminol chemiluminescence inhibition detection
Authors:Amir Waseem  Mohammad Yaqoob  Masood Ahmed Siddiqui
Affiliation:Department of Chemistry , University of Balochistan , Quetta, Pakistan
Abstract:A flow-injection procedure is described for the determination of carbaryl based on its inhibition effect on luminol-cobalt(II) chemiluminescence reaction in alkaline medium in the presence of hydrogen peroxide. The calibration data over the range 5.0?×?10?7 to 20?×?10?6?M give a correlation coefficient (r 2) of 0.9972 with relative standard deviations (RSD; n?=?4) in the range of 1.0–2.1% with a limit of detection (3?×?blank noise) of 2.37?×?10?7?M for carbaryl. The sample throughput was 120?h?1. The effects of some carbamates, anions, and cations were studied on luminol CL system for carbaryl determination. The proposed method has been applied to determine carbaryl in natural waters.
Keywords:Flow-injection analysis  Chemiluminescence  Inhibition  Carbaryl