首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Vermiculite clay mineral as an effective carbon paste electrode modifier for the preconcentration and voltammetric determination of Hg(II) and Ag(I) ions
Authors:Irena Grabec ?vegl  Mitja Kolar  Bo?idar Ogorevc  Boris Pihlar
Institution:Analytical Chemistry Laboratory, National Institute of Chemistry, P.O. Box 3040, 1001 Ljubljana, Slovenia, XX
Faculty of Chemistry and Chemical Technology, University of Ljubljana, Ljubljana, Slovenia, XX
Abstract:A vermiculite modified carbon paste electrode (VMCPE) was employed for the in situ preconcentration of traces of Hg(II) and Ag(I) via an ion-exchange route. Heavy metal ions were accumulated in Britton-Robinson (BR) buffer pH 7 for Hg(II) and pH 6 for Ag(I), and afterwards reduced at –0.7 V vs. Ag/AgCl in the separate measurement solution (BR buffer pH 5 + 0.05 mol/L NaNO3) prior to the anodic stripping square-wave voltammetric (ASSWV) detection. For Hg(II) ions, at 15 min accumulation, a linear range from 1.0 × 10–7 to 8.0 × 10–6 mol/L was obtained, with a 5.7 × 10–8 mol/L limit of detection. The VMCPE response was linear for Ag(I) ions in the concentration range from 2.0 × 10–7 to 8.0 × 10–6 mol/L, at 10 min accumulation with a corresponding limit of detection of 6.3 × 10–8 mol/L. The relative standard deviation of the analytical procedure including accumulation from a 5 × 10–7 mol/L solution of Hg (15 min) or Ag(I) (10 min), electrolysis, ASSWV detection, regeneration and activation of the VMCPE, was 4% (n = 6). The optimisation of the parameters for the application of the VMCPE in combination with ASSWV detection is presented and discussed.
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号