首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Development and validation of a liquid chromatography/tandem mass spectrometry method for the determination of the novel anticancer agent 17-DMAG in human plasma
Authors:Moreno-Farre Javier  Asad Yasmin  Pacey Simon  Workman Paul  Raynaud Florence I
Institution:Cancer Research UK Centre for Cancer Therapeutics, The Institute of Cancer Research, Haddow Laboratory, 15 Cotswold Road, Sutton, Surrey SM2 5NG, UK.
Abstract:An accurate, sensitive, robust and selective liquid chromatography/tandem mass spectrometry (LC/MS/MS) method for the determination of 17-(dimethylaminoethylamino)-17-demethoxygeldanamycin hydrochloride (17-DMAG) in human plasma has been developed and validated. Plasma samples were prepared by liquid/liquid extraction with ethyl acetate. The chromatographic separation was achieved within 9 min on a Synergy Polar column with a linear gradient and a mobile phase consisting of methanol and 0.1% formic acid in water. Detection of 17-DMAG and the internal standard (IS), olomoucine, was achieved by MS/MS with electrospray ionisation in positive ion mode. The calibration curve, ranging from 1.89 to 1890 nM, was linear r > 0.994 using a 1/y2 weighted linear regression. The assay showed no significant interferences from endogenous compounds. The lower limit of quantitation (LLOQ) was 1.89 nM, using 250 microL of plasma, with inter-assay precision (%RSD) and accuracy (%RE) values of 11.6% and -5.8%, respectively. Intra-assay precision ranged from 7.8-13.6%. The method described here is being used to evaluate the pharmacokinetic profiles of 17-DMAG given as a once weekly infusion in patients with advanced solid tumours.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号