Investigation of chemometric calibration performance based on different chemical matrix and signal-to-noise ratio. |
| |
Authors: | Soohwa Cho Hoeil Chung |
| |
Affiliation: | Department of Chemistry, College of Natural Sciences, Hanyang University, Haengdang-Dong, Seongdong-Gu, Seoul, Korea. |
| |
Abstract: | The chemometric calibration performance was systematically investigated by two parameters (changing the chemical matrix as well as the signal-to-noise ratio) of the NIR (near-infrared) spectrum. Three different analytes (hexane, cyclohexane, toluene) were selected and heptane was used as a solvent. The degree of spectral difference significantly affected the calibration performance. The largest structural difference between the analyte and the solvent provided the best calibration result for a given signal-to-noise ratio. Additionally, the signal-to-noise ratio of the spectra also directly influenced the calibration performance. Overall, the spectral difference and signal-to-noise ratio were the major factors for governing the chemometric calibration performance, especially in the low-concentration range. |
| |
Keywords: | |
|
|