首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Analysis of phenolic xenoestrogens by pressurized CEC with amperometric detection
Authors:Weimin Wu  Xiaomei Yuan  Xiaoping Wu  Xucong Lin  Zenghong Xie
Institution:1. Institute of Food Safety and Environmental Monitoring, Fuzhou University, P. R. China;2. Xiamen Huaxia Vocational College, Xiamen, P. R. China
Abstract:A new method, pressurized CEC with end‐column amperometric detection using carbon paste electrode, has been developed for the separation and determination of five phenolic xenoestrogens in chicken eggs and milk powder samples. Efficient separation of five analytes was performed by pressurized CEC using a mobile phase consisting of 60% v/v ACN and 40% v/v Tris buffer (5 mmol/L, pH 8.0), +6 kV of applied voltage and 7.0 MPa of supplementary pressure. Detection limits of 50, 5, 2, 10 and 20 ng/mL for pentachlorophenol, bisphenol‐A, 2,4‐dichlorophenol, 4‐tert‐octylphenol and 4‐nonylphenol, respectively, were achieved using carbon paste electrode as working electrode and +0.8 V as detection potential. Matrix solid phase dispersion extraction method had been employed during sample preparation procedure, and mean recoveries ranged from 79.2 to 102.6% at different concentrations of phenolic xenoestrogens for spiked egg and milk powder samples were obtained.
Keywords:Amperometric detection  Phenolic xenoestrogens  Pressurized CEC
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号