Preconcentration of trace elements by partial dissolution of the matrix—multielement preconcentration from manganese |
| |
Authors: | E. Jackwerth J. Messerschmidt R. Höhn |
| |
Affiliation: | Institut für Spektrochemie und angewandte Spektroskopie, Bunsen-Kirchhoff-Str. 11, 4600 Dortmund 1, Federal Republic of Germany |
| |
Abstract: | ![]() Trace elements such as Ag, Au, Bi, Cd, Co, Cu, Fe, Ga, In, Ni, Pb, Pd and Tl, can be preconcentrated with recoveries of better than 95% from high-purity manganese if the sample is coated with a thin layer of mercury before its dissolution in HCl to a small residue. For determination of the trace elements, the residue is completely dissolved in aqua regia. After separation of the mercury by reductive precipitation, Bi, Cd, Co, Cu, Fe, Ga, In, Ni, Pb, and Tl are determined by flame a.a.s. ('injection method'). Ag, Au and Pd are determined in the mercury-containing solution. The relative standard deviation was usually about 5%; Cu, Fe and Pb were proved to be inhomogeneously distributed in the sample (electrolytically produced manganese), and so the standard deviations were considerably greater. The detection limits for the different elements were between 0.6 and 0.004 ppm, depending on the sensitivity of their a.a.s. determination. The theoretical basis of this preconcentration method and its applicability to multielement analysis of different high-purity metals are discussed. |
| |
Keywords: | |
本文献已被 ScienceDirect 等数据库收录! |
|