首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Enhanced Performance of CNT‐doped Imine Based Receptors as Fe(III) Sensor Using Potentiometry and Voltammetry
Authors:Susheel K&#x;Mittal  Sanjeev Kumar  Navneet Kaur
Institution:Susheel K.?Mittal,Sanjeev Kumar,Navneet Kaur
Abstract:A highly sensitive and selective potentiometric and voltammteric assay for the detection of Fe3+ using (E)‐3‐((2‐(2‐(2‐aminoethylamino) ethylamino) ethylimino)methyl)‐4H‐chromen‐4‐one (IFE(III)) ionophore was developed. To demonstrate the ion‐to‐electron transfer ability of MWCNT, these were incorporated in the ion‐selective membrane and response characteristics of Fe3+ electrode was compared with those of the traditional ion selective electrode. The electrode showed an improved Nernstian slope, lower detection limit, response time of less than 5 s and working in a pH range of 3.0 to 8.0. Differential pulse voltammetric studies were performed for IFE(III)‐Fe3+ complex in DMSO solvent medium at glassy carbon (GC) electrode. A linear relationship between the cathodic peak current and concentration of Fe3+ was observed in the range of 1.6×10?5 to 4.4×10?5 mol/L with a detection limit of 5.2×10?8 mol/L. The electrode shows remarkable selectivity for Fe3+ ions over alkali, alkaline earth, transition and heavy metal ions. The optimized electrode was successfully applied for the determination of Fe3+ ion in different real‐life samples using potentiometric technique. Theoretical calculations were used to support the complexation behavior of Fe3+ with IFE(III).
Keywords:Schiff base Ionophore  Potentiometric sensor  Voltammetric sensor  MWCNT effect on ISE  Fe(III) determination
设为首页 | 免责声明 | 关于勤云 | 加入收藏