Determination of (R)- and (S)-phenprocoumon in human plasma by enantioselective liquid chromatography/electrospray ionisation tandem mass spectrometry |
| |
Authors: | Kammerer Bernd Kahlich Rainer Ufer Mike Laufer Stefan Gleiter Christoph H |
| |
Affiliation: | Institute of Pharmacology and Toxicology, Department of Clinical Pharmacology, University Hospital Tübingen, 72076 Tübingen, Germany. Bernd.Kammerer@uni-tuebingen.de |
| |
Abstract: | Phenprocoumon is a commonly used oral anticoagulant of the coumarin type, and has found extensive clinical use in the treatment of thrombophlebitis, pulmonary embolism and atrial fibrillation. In the course of a clinical study to investigate the influence of genetic polymorphisms of the CYP2C9 enzyme on phenprocoumon metabolism, we developed a new enantioselective liquid chromatography/electrospray ionisation tandem mass spectrometry (LC/MS/MS) method to quantify (R)- and (S)-phenprocoumon in human plasma. HPLC separation of the enantiomers was achieved on a Chira-Grom-2 column under isocratic conditions using a water/acetonitrile/formic acid eluent. For detection and quantification a triple-quadrupole MS system was used in the selected reaction monitoring (SRM) mode. As an internal standard the structurally homologous compound warfarin was chosen. The detector response was linear with a correlation coefficient of 0.988-0.999 for (R)-phenprocoumon and 0.989-0.999 for (S)-phenprocoumon in the investigated concentration range between 62.5 and 1000 ng/mL (per enantiomer). The limit of detection (LOD) was 12.5 ng/mL. |
| |
Keywords: | |
本文献已被 PubMed 等数据库收录! |
|