New approach to calibrating conductivity meters in the low conductivity range
Authors:Petra Spitzer  Bruno Rossi  Yves Gaignet  Stéphane Mabic  Uwe Sudmeier
Institution:(1) Physikalisch-Technische Bundesanstalt, Metrology in Chemistry, Bundesallee 100, 38116 Braunschweig, Germany;(2) Lab Water Division, Millipore, St Quentin-Yvelines, France
Abstract:There is currently a major issue with the calibration of conductivity meters used for high purity water: the lack of availability of a reference material or reference methods for low conductivity ranges (conductivity below 1 mgrS cm–1 at 25.0 °C, resistivity >1 MOHgr cm at 25.0 °C). This paper describes the current status of conductivity measurements in high purity water. A new and improved approach, currently being investigated, should allow us to make the calibration of conductivity meters used for low conductivity ranges traceable to the SI.Milipore, Milli-Q and Elix are registered trademarks of Millipore Corporation.
Keywords:Electrolytic conductivity  High purity water  Calibration  Primary methods  Metrology in chemistry