首页 | 本学科首页   官方微博 | 高级检索  
     


Micromachined array tip for multifocus fiber-based optical coherence tomography
Authors:Yang Victor X D  Munce Nigel  Pekar Julius  Gordon Maggie L  Lo Stewart  Marcon Norman E  Wilson Brian C  Vitkin I Alex
Affiliation:Department of Medical Biophysics, University of Toronto, Toronto, Ontario M5G 2M9, Canada. victor.yang@utoronto.ca
Abstract:
High-resolution optical coherence tomography demands a large detector bandwidth and a high numerical aperture for real-time imaging, which is difficult to achieve over a large imaging depth. To resolve these conflicting requirements we propose a novel multifocus fiber-based optical coherence tomography system with a micromachined array tip. We demonstrate the fabrication of a prototype four-channel tip that maintains a 9-14-microm spot diameter with more than 500 microm of imaging depth. Images of a resolution target and a human tooth were obtained with this tip by use of a four-channel cascaded Michelson fiber-optic interferometer, scanned simultaneously at 8 kHz with geometric power distribution across the four channels.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏