Simple high-performance liquid chromatographic method for the determination of captopril in biological fluids. |
| |
Authors: | S Gao W Tian S Wang |
| |
Affiliation: | Department of Clinical Pharmacology, Changhai Hospital, Shanghai, China. |
| |
Abstract: | A rapid, simple and sensitive column-switching high-performance liquid chromatographic procedure for the determination of captopril in plasma and urine had been developed. p-Bromophenacyl bromide was used as a derivatizing reagent to react with captopril to form a product that showed ultraviolet-absorbing properties. For plasma samples the protein was removed with 6% perchloric acid before injection. The urine samples were directly injected into the chromatograph. The column-switching system was equipped with a pre-column (5 cm x 0.5 cm I.D.) packed with muBondapak C18 (37-50 microns) and an analytical column (15 cm x 0.5 cm I.D.) packed with YWG-C18, 10 microns. Impurities were washed from the pre-column with 0.2% acetic acid and the retained substances were eluted into the analytical column with acetonitrile-water-acetic acid (35:65:0.4, v/v). Captopril was detected at 260 nm. The calibration curve was linear in the range 20-1000 ng/ml for plasma and 10-200 micrograms/ml for urine. The recoveries averaged 103.2 and 99.5% for plasma and urine, respectively. The coefficients of variation were all less than 10%. |
| |
Keywords: | |
|
|