Direct determination of hydrogen cyanide in cigarette mainstream smoke by ion chromatography with pulsed amperometric detection |
| |
Authors: | Zhang Zi-Wei Xu Ying-Bo Wang Cheng-Hui Chen Kai-Bo Tong Hong-Wu Liu Shao-Min |
| |
Affiliation: | Department of Chemistry, University of Science and Technology of China, Hefei 230026, China. |
| |
Abstract: | The determination of hydrogen cyanide in cigarette mainstream smoke has been achieved by ion chromatography (IC) with pulsed amperometric detection (PAD). The proposed method of totally trapping whole cigarette mainstream smoke by Cambridge filters, which are treated with sodium hydroxide/ethanol solution, possesses the advantage of fast analysis time over the widespread used solution absorption method. The possible co-existing interferents are evaluated under the optimized detection conditions and excellent recoveries of cyanide are obtained. The cyanide content of absorption solution can be directly determined by the optimized IC-PAD method without any pretreatments. The linear range is 0.0147-2.45 μg/mL with R2 value of 0.9997. The limit of the detection is 3 μg/L for a 25 μL injection loop. The overall relative standard deviation of the method is less than 5.20% and the recovery range from 94.3% to 101.0%. The results obtained from the developed method are in good agreement with that of continuous flow analyzer (CFA) method. |
| |
Keywords: | Hydrogen cyanide Ion chromatography Pulsed amperometric detection Cigarette Smoke trapping |
本文献已被 ScienceDirect PubMed 等数据库收录! |