首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Determination of total iron in ground waters and multivitamin tablets using a solid-phase reactor with tiron immobilised on amberlite ion-exchange resin in a flow injection system
Authors:J F van Staden and L G Kluever
Institution:(1) Department of Chemistry, University of Pretoria, Pretoria 0002, South Africa e-mail:JvStaden@Scientia.up.ac.za, ZA
Abstract:A method is described for the flow injection determination of total iron as Fe(III) using a solid-phase reactor containing disodium-1,2-dihydroxybenzene-3,5-disulphonate (tiron) as substrate. The iron(III) reacted with tiron to form a complex which absorbs strongly at 667 nm, where it was measured spectrophotometrically. The system has a linear range of 1 to 50 mg L–1 with a detection limit of 0.67 mg L–1. It is suitable for the determination of total iron in multivitamin tablets and iron-rich ground waters, with a relative standard deviation of better than 1.1%. The results obtained compared favourably with the certified values and a standard ICP-AES method. Received: 12 November 1997 / Revised: 9 March 1998 / Accepted: 15 March 1998
Keywords:
本文献已被 SpringerLink 等数据库收录!