Preconcentration and voltammetric determination of trace myoglobin at a 6-mercaptopurine modified silver electrode |
| |
Authors: | Y.-T. Long Jun-Jie Zhu H.-Y. Chen |
| |
Affiliation: | (1) Department of Chemistry, The State Key Laboratory of Coordination Chemistry, Nanjing University, Nanjing 210093, P.R.China e-mail: HYChen@public1.ptt.js.cn, CN |
| |
Abstract: | 6-Mercaptopurine(6-MP) is covalently modified onto a silver electrode to construct a chemically modified electrode (CME). It exhibits the capability of selectively complexing myoglobin and can be applied as a selective biosensor for this compound. Myoglobin is accumulated onto the CME at 0.32 V (vs. SCE); after exchanging the medium it is determined by differential pulse stripping voltammetry. A cathodic stripping peak is obtained at 0.15 V (vs. SCE) by scanning the potential from 0.35 to –0.1 V. The peak currents are linearly proportional to the concentration of myoglobin in the range of 0.2 ∼ 4 μg/mL with a relative standard deviation of 7.8%. The detection limit is about 0.1 μg/mL. The mechanism of the complexation is also discussed. Received: 30 January 1997 / Revised: 7 May 1996 / Accepted: 10 May 1996 |
| |