首页 | 本学科首页   官方微博 | 高级检索  
     


Advanced pulse calibration techniques for the quantitative analysis of TG–FTIR data
Authors:Katia Marsanich   Federica Barontini   Valerio Cozzani  Luigi Petarca
Affiliation:

a Dipartimento di Ingegneria Chimica, Chimica Industriale e Scienza dei Materiali, Università degli Studi di Pisa, via Diotisalvi n.2, 56126 Pisa, Italy

b Gruppo Nazionale per la Difesa dai Rischi Chimico-Industriali ed Ecologici, Consiglio Nazionale delle Ricerche, via Diotisalvi n.2, 56126 Pisa, Italy

Abstract:
Two different pulse calibration techniques to estimate the total quantities of evolved gaseous substances formed in thermogravimetric (TG)–FTIR runs were compared and assessed. A gas-pulse calibration method was based on the use of a specific device able of sending a known quantity of a gaseous compound of interest to the FTIR analyzer. A second calibration method was based on the vaporization in the TG analyzer of liquid solutions of the compound of interest. Data obtained by these techniques were compared to those from conventional concentration-based calibration. The results confirmed the reliability of pulse calibration techniques to obtain quantitative data on evolved gaseous products in TG–FTIR applications. Moreover, both the gas-pulse and the vaporization-based calibration techniques proved to have several advantages with respect to conventional techniques. Among these are the need of a more limited number of standards and no need for online gas dilution systems.
Keywords:TG–FTIR   Hyphenated techniques   Quantitative evolved gas analysis   Calibration techniques   Thermal analysis
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏