首页 | 本学科首页   官方微博 | 高级检索  
     


Determination of Lead(II), Copper(II), Zinc(II), Cobalt(II), Cadmium(II), Iron(III), Mercury(II) using sequential injection extractions
Authors:van Staden J F  Taljaard R E
Affiliation:

Department of Chemistry, University of Pretoria, Pretoria 0002, South Africa

Abstract:Sequential injection analysis is still dominated by single component analysis. In this proposed robust, economical simple instrumental system seven different metal ions are determined simultaneously using thin-film sequential injection extraction (SIE) with multivariate calibration and multiwavelength detection. Dithizone, in ethanol, is used as extractant and the metal dithizonates’ spectra are generated by a diode array spectrophotometer between 300 and 700 nm. The SI thin-film extraction using water miscible with ethanol works due to the hydrophobic interaction of ethanol with the Teflon wall to create a thin film. A sample frequency of 27 samples per hour was obtained with a sample carry-over of less than 1%. The results of the proposed sequential injection extraction system compare favourably with the results obtained by using standard atomic absorption spectrometry (AAS) methods on conventional extraction samples.
Keywords:Sequential injection extractions   SI thin-film extraction   Diode array spectrophotometer
本文献已被 ScienceDirect PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏