Simultaneous quantitative analysis of ketanserin and ketanserinol in plasma by RP-HPLC with fluorescence detection
Authors:Yassen A  Hanff L M  Vulto A G
Institution:Department of Hospital Pharmacy, Erasmus MC Rotterdam, The Netherlands.
Abstract:A sensitive and selective high-performance liquid chromatographic assay for the quantification of ketanserin and ketanserinol in human plasma was developed and validated. The procedure involves extraction of ketanserin and ketanserinol from plasma using an Extrelut NT-1 solid-phase extraction column. The chromatograph was equipped with a Hypersil BDS column (100 x 4.5 mm, 3 micro m particle size). Separation was performed with a mixture of acetate buffer 0.01 M, pH 4.9-methanol-acetonitrile (52:40:8, v/v/v). Detection was performed with fluorescence detection (lambda(ex) = 332 nm and lambda(em) = 410 nm). Calibration curves were linear (r(2) = 0.999) in the range 0-400 ng/mL for both ketanserin and ketanserinol. The repeatability coefficient for ketanserin and ketanserinol was 3.1 and 3.0%, respectively. The reproducibility coefficient for ketanserin and ketanserinol was 10.5 and 9.1%, respectively. The limit of quantification for both ketanserin and ketanserinol was 2.0 ng/mL. The mean recovery yield for both ketanserin and ketanserinol was 60%. In an 8 h work day approximately 60 samples, including calibration and reference standards, could be processed.
Keywords:pregnancy  pre‐eclamptic  dosage regimens  neonate plasma level
本文献已被 PubMed 等数据库收录!