Frequency stability of an optically pumped cesium beam frequency standard |
| |
Authors: | V. Candelier V. Giordano A. Hamel G. Théobald P. Cérez C. Audoin |
| |
Affiliation: | (1) Laboratoire de l'Horloge Atomique, Unité Propre de Recherche du CNRS associée à l'Université Paris-Sud, Bât. 221, Université Paris-Sud, F-91405 Orsay Cedex, France;(2) Present address: CEPE-44 avenue de la Glacière, F-95100 Argenteuil, France |
| |
Abstract: | Results obtained in an experimental optically pumped cesium beam frequency standard in which a single semiconductor laser is used for the state selection and the atom detection are reported. The separation between the two interaction regions is equal to 21 cm. This gives a 500 Hz linewidth which is observed with a signal to noise ratio equal to 10,000 in a 1 Hz noise bandwidth. A quartz crystal oscillator is frequency controlled by the atomic transition. The measured short term frequency stability is given by y( )=2×10–12 –1/2 for 1 s< 500 s. Prospects for improvement of this frequency stability are discussed. |
| |
Keywords: | 32.80B 35.10F 35.80 |
本文献已被 SpringerLink 等数据库收录! |
|