首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Determination of benzalkonium chloride in eye care products by high-performance liquid chromatography and solid-phase extraction or on-line column switching.
Authors:L Elrod  T G Golich  J A Morley
Institution:PPD Physical Analytical Chemistry Department, Abbott Laboratories, North Chicago, IL 60064.
Abstract:Benzalkonium chloride (BAK) is a mixture of alkylbenzyldimethylammonium chlorides, which is commonly used as a bacteriostat. In this work, the three major homologues of BAK are quantitated in the over-the-counter eye care products Murine and Murine Plus using high-performance liquid chromatography (HPLC). The analytes are separated from various product excipients and concentrated by either solid-phase extraction onto Sep-Pak C18 cartridges or by an on-line column-switching technique using 1-cm reversed-phase precolumns. Absolute recoveries of BAK homologues by the solid-phase extraction technique ranged from 97.2 to 98.7% for standards and from 98.0 to 98.4% for samples. Absolute recovery of the BAK homologues by the column-switching technique was 101.3% for standards and ranged from 99.9 to 103.7% for samples. Relative recoveries were quantitative by both techniques. Assay precision (R.S.D. values) were +/- 2.2% to +/- 2.6% and +/- 0.4% to +/- 0.8% by solid-phase extraction and column-switching techniques, respectively. The method provides advantages of high sample throughput, excellent column life and automation.
Keywords:
设为首页 | 免责声明 | 关于勤云 | 加入收藏