Automated optimization of split ratio in capillary gas chromatography
Authors:Y Nishikawa  S Miyoshi  M Shyoji  G Braun
Abstract:An electronic gas-flow controller system called Advanced Flow Control (AFC), which controls not only carrier gas pressure at the column inlet but also the total gas flow including the split flow, was designed and evaluated. BASIC programming of repetitive analyses of standard mixtures under varied split ratios and pressure programs allowed automated optimization of those conditions for the desired column loading and resolution between adjacent peaks.
Keywords:Capillary gas chromatography  Electronic flow controller  Advanced flow controller (AFC)  Pressure programming  Flow programming  Automatic optimization  Split injection