首页 | 本学科首页   官方微博 | 高级检索  
     


Full energy peak efficiency calibration of HPGe detector for point and extended sources using Monte Carlo code
Authors:Chhavi Agarwal  Sanhita Chaudhury  A. Goswami  M. Gathibandhe
Affiliation:(1) Radiochemistry Division, Bhabha Atomic Research Centre, Mumbai, 400085, India;(2) Reactor Projects Division, Bhabha Atomic Research Centre, Mumbai, 400085, India;
Abstract:
Monte Carlo simulation is important to get efficiencies for cases where the experimental efficiencies are difficult to get such as for samples with nonstandard geometries and for large samples. In this paper, efficiency of the HPGe detector, routinely used in our lab for a variety of samples, has been computed for point source geometry and its parameters has been optimized to match MCNP and experimental efficiencies within 5% at different sample to detector distances. This optimized geometry was then validated by efficiency transfer to other geometries.
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏