首页 | 本学科首页   官方微博 | 高级检索  
     检索      


On-line trace preconcentration / matrix separation by high-performance flow atomic spectrometry (HPF-FLAME AAS)
Authors:H Berndt  A Müller and G Schaldach
Institution:(1) Institut für Spektrochemie und angewandte Spektroskopie, Bunsen-Kirchhoff-Strasse 11, W-4600 Dortmund 1, Germany
Abstract:Summary In HPF-atomic spectrometry a high-performance flow / hydraulic high-pressure nebulization (HPF / HHPN) system is used for sample introduction and aerosol generation. By employment of techniques common in HPLC or ion chromatography, on-line trace element preconcentration / matrix separation and atomic spectrometric trace determinations can be carried out. Preconcentration of trace elements in samples of drinking water allows determinations within the lower mgrg/L region by using flame AAS. On-line trace element preconcentration / matrix separation from aluminium leads to detection limits of approx. 0.1 to 1 mgrg/g within less than 3 min of total analysis time. Dependent on concentration and the element involved, the relative standard deviation amounts to approx. 2 to 4% (2.5 to 25 mgrg/g traces/aluminium).Dedicated to Professor Dr. Wilhelm Fresenius on the occasion of his 80th birthday
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏