首页 | 本学科首页   官方微博 | 高级检索  
     


On-line microwave sample pretreatment for the determination of mercury in blood by flow injection cold vapor atomic absorption spectrometry
Authors:Guo T  Baasner J
Affiliation:

Atomic Absorption Product Department, Bodenseewerk Perkin-Elmer GmbH Postfach 101164 W-7770 Überlingen, Germany

Abstract:
A method for on-line treatment of whole blood in a microwave oven and determination of mercury by flow injection cold vapor atomic absorption spectrometry was developed. After dilution of the whole blood and addition of oxidant, all further treatment and measurement were performed automatically, on-line. Recoveries of five mercury compounds were complete. Good agreement between measured and recommended values of mercury in whole blood reference materials was obtained. Measured mercury values also agreed with results from other accepted methods. Sample throughput was about 45 measurements/hr. Detection limit (3s) in diluted sample was 0.1 μg/l corresponding to 1μg/l Hg in whole blood. The RSD value at 0.5 μg/l Hg in the diluted sample was 6–7% (11 measurements and 0.5 ml sample volume). Mercury concentrations between 1 and 150 μg/l in whole blood can be measured using this method. For three replicate measurements, 0.5 ml of whole blood is required.
Keywords:
本文献已被 ScienceDirect PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号