首页 | 本学科首页   官方微博 | 高级检索  
     


A reversed-phase high-performance liquid chromatographic method to analyze retinal isomers
Authors:Cia D  Bonhomme B  Azim M  Wada A  Doly M  Azaïs-Braesco W
Affiliation:Laboratoire de Biophysique Sensorielle, Facultés de Médecine et de Pharmacie, Clermont-Ferrand, France. david.cia@u-clermont1.fr
Abstract:A high-performance liquid chromatographic (HPLC) procedure was developed to separate all-trans-, 13-cis-, 11-cis- and 9-cis-retinal isomers. Two reversed-phase Vydac C18 columns in series were used with an isocratic solvent system of 0.1 M ammonium acetate-acetonitrile (40:60, v/v) as mobile phase and all-trans-9-(4-methoxy-2,3,6-trimethylphenyl)-3,7-dimethyl-2,4,6,8-no natetraene-1-ol (TMMP) as internal standard. Prior to HPLC, the retinal isomers were efficiently extracted in their original isomeric conformation using dichloromethane-n-hexane in the presence of formaldehyde. This technique is suitable for the assay of 11-cis- and all-trans-retinal isomers in retina.
Keywords:
本文献已被 PubMed 等数据库收录!