Abstract:Rapamycin is a potent immunosuppressive drug capable of significantly reducing acute graft rejection in kidney, liver and heart transplant patients. Its immunosuppressive activity and adverse effects have been related to rapamycin concentration, and therapeutic drug monitoring of the drug is deemed appropriate. This work was aimed at developing a new quantification method based on the isolation of the M+Na]+ ion as precursor and its further fragmentation through an ion trap mass spectrometer equipped with an electrospray ionization source. A limit of detection (LOD) of 0.7 ng/mL was obtained, while the lower limit of quantification (LLOQ) was 2.4 ng/mL. The accuracy and reproducibility of the responses were evaluated and compared with results obtained when the M+NH4]+ ion was chosen as the precursor in a triple quadrupole mass spectrometer. In this case the LOD was 0.5 ng/mL and the LLOQ 1.7 ng/mL. Data showed that it would be possible to use the quantification of the sodiated species for the routine determination of rapamycin, as an alternative to the commonly adopted method based on the ammoniated complex.