Use of non-segmented flow,post-columns reaction detection with miniaturized HPLC systems |
| |
Authors: | J. A. Apffel U. A. Th. Brinkman R. W. Frei |
| |
Affiliation: | (1) Department of Analytical Chemistry, Free University, De Boelelaan 1083, 1081 HV Amsterdam, The Netherlands |
| |
Abstract: | Summary The use of non-segmented flow, post-column reaction detection is evaluated for use with miniaturized HPLC. Non-segmented open-tubular reactors with internal diameters 0.1 mm to 0.25 mm and packed-bed reactors with internal diameters of 1 mm, filled with 5 and 10 m particles are evaluated theoretically and experimentally with respect to band broadening and pressure drop characteristics. An integrated system consisting of miniaturized HPLC (1 mm i. d.) columns and compatible hardware is described. An example of the separation of catecholamines is given. |
| |
Keywords: | Miniaturized HPLC Post-column reaction detection Catecholamines Packed-bed reactors Non-segmented open-tubular reactors |
本文献已被 SpringerLink 等数据库收录! |