首页 | 本学科首页   官方微博 | 高级检索  
     


Automated trace level determination of glyphosate and aminomethyl phosphonic acid in water by on-line anion-exchange solid-phase extraction followed by cation-exchange liquid chromatography and post-column derivatization.
Authors:J Patsias  A Papadopoulou  E Papadopoulou-Mourkidou
Affiliation:Pesticides Science Laboratory, Aristotle University of Thessaloniki, Greece.
Abstract:
An automated method based on the on-line coupling of anion-exchange solid-phase extraction (SPE) and cation-exchange liquid chromatography followed by post-column derivatization and fluorescence detection has been developed for the trace level determination of glyphosate and its primary conversion product aminomethyl phosphonic acid (AMPA) in water. PRP-X100 poly(styrene-divinylbenzene)-trimethylammonium anion-exchange cartridges (20 x 2 mm, 10 microm) were selected for the SPE of glyphosate and AMPA. The ionic compounds present in the samples strongly influenced the extraction of both analytes; however, when an on-line ion-exchange clean-up step was introduced before sample SPE, the problem was largely solved. By processing 100-ml samples detection limits better than 0.02 microg/l for glyphosate and 0.1 microg/l for AMPA were achieved in river water. Both analytes were unstable in solution and the approach of storing samples on the PRP-X100 SPE cartridges was evaluated for a period of 1 month under three different storage conditions (deep freeze, refrigeration and 20 degrees C).
Keywords:
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号