Electron capture detector performance parameters in view of the response presentation forms |
| |
Authors: | P. Rotocki J. Lasa |
| |
Affiliation: | (1) Institute of Nuclear Physics, Radzikowskiego Street 152, 31-342 Cracow, Poland |
| |
Abstract: | Summary Four most frequently used presentation forms of the ECD response are described. Equations describing the dependence of the sensitivity, linearity and minimum detectable concentration of the ECD working under constant frequency mode of operation on the pulse period are derived from Wenthworth’s kinetic model for all four forms mentioned above. The forms presented are compared on the basis of equations derived. Results of the calculations are given and some predictions are experimentally tested for sulphur hexafluoride. Good agreement was found between the predicted and observed relationships. Presented at the 15th International Symposium on Chromatography, Nürnberg, October 1984 |
| |
Keywords: | Electron capture detection Constant frequency ECD ECD performance parameters ECD response presentation |