Selenium speciation by high-performance anion-exchange chromatography-post-column UV irradiation coupled with atomic fluorescence spectrometry |
| |
Authors: | Liang Lina Mo Shumin Zhang Ping Cai Yaqi Mou Shifen Jiang Guibin Wen Meijuan |
| |
Affiliation: | State Key Laboratory of Environmental Chemistry and Ecotoxicology, Research Center for Eco-Environmental Sciences, Chinese Academy of Sciences, P.O. Box 2871, Beijing 100085, China. |
| |
Abstract: | A technique for the speciation of selenomethylcysteine (SeMeCys), selenocystine (SeCys), selenite [Se(IV)] and selenomethionine (SeMet) was established in this paper using high-performance anion-exchange chromatography coupled with atomic fluorescence spectrometry (HPAEC-AFS). Analytes were separated on an AminoPac PA10 column and then digested by on-line ultraviolet (UV) irradiation, which destroyed organic compound structure. Hydride generation was used as an available sample introduction technique for atomic fluorescence detection. The detection limits of four compounds were 1-5 microg/L (250 microL injection, 10 times of the baseline noise). The relative standard deviations (RSDs), calculated from seven consecutive injections of 100 microg/L standard mixtures, were from 2 to 4%. Selenious yeast tablet, which had been proposed as selenium supplement, and human urine collected from a volunteer were analyzed. Good spiked recoveries from 86 to 103% were obtained. |
| |
Keywords: | |