首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Studying a kind of portable ultra-bright microfocus x-raysource
Authors:Wang Kai-Ge  Wang Lei and Niu Han-Ben
Institution:Institute of Photonics & Photonic Technology, Northwest University, Xi'an 710069, China; Institute of Optoelectronics, Shenzhen University, Shenzhen 518060, China; Xi'an Institute of Applied Optics, Xi'an 710065, China
Abstract:This paper studies the properties of a kind of portable ultra-bright microfocus x-ray source with the Monte-Carlo method in detail. The new x-ray source consists of an electron-emission system, an electrostatic focusing system and a metal target. A crystal Lanthanum Hexaboride cathode, a Wehnelt grid and an extracted electrode compose the triode electrode electron-gun system. Two equal radius cylinder electrodes form the focusing system. The key factors determining the focus properties of the electron beam such as the ratio $D_{\rm w}/H$, grid bias $V_{\rm g}$, and the properties of the extracted electrode are numerically studied. The calculated results reveal that when $D_{\rm w}$/$H$, $V_{\rm g}$, the length of the extracted electrode, and the distance between the grid and the extracted electrode equals 5, --0.6~kV, 10~mm, and 8~mm respectively, the electron beam focal spot can be concentrated down to 9~$\mu $m in radius and a reasonable focal length about 72.5~mm can be achieved, at the same time, the cathode emission currents can be as high as 30~mA.
Keywords:microfocus x-ray source  high brightness  micro-focal spot  Monte-Carlo method
本文献已被 维普 等数据库收录!
点击此处可从《中国物理 B》浏览原始摘要信息
点击此处可从《中国物理 B》下载免费的PDF全文
设为首页 | 免责声明 | 关于勤云 | 加入收藏