Evaluation of on-line coupling size exclusion chromatography electrothermal atomic absorption spectrometry for selenium speciation |
| |
Authors: | Francisco Laborda María V. Vicente José M. Mir J. R. Castillo |
| |
Affiliation: | (1) Analytical Spectroscopy and Sensors Group (G.E.A.S.), Department of Analytical Chemistry, University of Zaragoza, E-50009 Zaragoza, Spain, ES |
| |
Abstract: | Chromatographic effluents were on-line analyzed by Zeeman-ETAAS, using a flow-through cell placed in a graphite furnace autosampler as interface. To obtain high sampling rates, the use of fast graphite furnace programmes was studied. Conventional programmes of 96 s were reduced to 18 s by using hot injection (120 °C) and reducing the charring step to 2 s. The increase of the injection volume from 20 to 60 μl lengthened the programme to 46 s. Nickel had to be added to get a comparable response for both inorganic and organic selenium species (selenite and selenomethionine) and to reduce the interferent effect produced in presence of the chromatographic eluent (TRIS 0.01 mol l-1, NH4NO3 0.1 mol l-1, pH 7). The optimized conditions were applied to the speciation of selenium in human erythrocyte lysates by size exclusion LC-ETAAS. Using a high performance size exclusion column selenium could be assigned to proteins of 100 and 35 kDa. Detection limits in the range of 1 ng (2 μg l-1 for 500 μl injection volume) were obtained for the combined technique. Received: 9 October 1996/Revised: 8 July 1996/Accepted: 14 July 1996 |
| |
Keywords: | |
本文献已被 SpringerLink 等数据库收录! |
|