CR-39 and Lexan calibrated as low-LET radiation dosimeter, for three Mexican irradiation facilities
Authors:L. Tavera, M. Balc  zar, H. Matamoros,H. Carrasco
Affiliation:

aInstituto Mexicano del Petróleo, Eje Central Lázaro Cárdenas No 152, México DF 07730, México

bInstituto Nacional de Investigaciones Nucleares, Apartado Postal 18-1027, México DF 11801, México

Abstract:An experimental calibration of two common nuclear track detectors, CR-39 and Lexan, for gamma and electrons radiation, was performed using various irradiation facilities. The dose response was obtained as a function of two parameters, the bulk etch rate and the UV absorbance for a wide dose range from 10 to 1000 kGy. The bulk etch rate sensitivity, for gamma and electrons, in CR-39 detector is higher than for Lexan detector. Lexan has a well-defined UV absorbance spectrum, but presents saturation for doses higher than 500 kGy, the same saturation characteristic is observed for the corresponding bulk etch rate response. For electron and gamma radiation, CR-39 shows a good response for doses from 10 kGy up to 1000 kGy, where data fit well an exponential curve for electrons and a lineal curve for gamma radiation.
Keywords:Gamma dosimeter   Electron dosimeter   CR-39   Lexan