首页 | 本学科首页   官方微博 | 高级检索  
     


ConFlo III - an interface for high precision delta(13)C and delta(15)N analysis with an extended dynamic range
Authors:RA Werner  BA Bruch  WA Brand
Affiliation:Max-Planck-Institut für Biogeochemie, P.O. Box 10 01 64, 07701 Jena, Germany.
Abstract:A newly developed interface coupling a CHN combustion device (elemental analyser 'EA') to an isotope ratio mass spectrometer is described and evaluated. The purpose of the device is to extend the dynamic range of delta(13)C and delta(15)N analysis from less than 2 orders of magnitude to more than 3 orders of magnitude. Carbon isotope ratio measurements of atropine as a model compound have been performed analysing between 1 μg to 5 mg C with acceptable to excellent precision (0.6 to 0.06 per thousand, delta-notation). The correction due to the blank signal is critical for sample amounts smaller than 4 μg C. The maximum sample weight is determined by the combustion capacity of the EA. Larger sample amounts are measured using dilution of a small part of the EA effluent with helium. The dilution mechanism works virtually free of isotope fractionation. Copyright 1999 John Wiley & Sons, Ltd.
Keywords:
本文献已被 PubMed 等数据库收录!