A comparative study of myo-inositol quantification using LCmodel at 1.5 T and 3.0 T with 3 D 1H proton spectroscopic imaging of the human brain
Authors:Srinivasan Radhika  Vigneron Daniel  Sailasuta Napapon  Hurd Ralph  Nelson Sarah
Affiliation:Magnetic Resonance Science Center, University of California, AC109, Box 1290, San Francisco, CA 94143, USA. rsriniva@mrsc.ucsf.edu
Abstract:Myo-inositol is a strongly coupled system and resonates at four chemical shift positions. At 1.5 T, only the singlet component at 3.57 ppm is detected. However, at 3 T this resonance is resolved into its components at 3.55 ppm and 3.61 ppm. Due to the increased spectral resolution and signal-to-noise ratio, it is anticipated that the quantification of myo-inositol should improve at 3 T. Using data from normal controls and the LCmodel quantification procedure, we found that the quantification precision, reproducibility and detection sensitivity of myo-inositol is significantly better at 3 T relative to 1.5 T.