首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Establishment of traceability of ammonium nitrogen determination in wastewater
Authors:Email author" target="_blank">Andreja?DrolcEmail author  Milenko?Ro?  Magda?Cotman
Institution:(1) National Institute of Chemistry, Hajdrihova 19, P.O. Box 660 , 1000 Ljubljana, Slovenia
Abstract:A case study is presented for the establishment of traceability for ammonium nitrogen determination in wastewater in a routine laboratory in order to fulfil the requirements of ISO/IEC standard 17025. The necessary relevant information was obtained from the method validation data, the quality control data and equipment calibration certificates. The method of measurement is described together with the measurement equation, selected traceable reference standards and the associated measurement uncertainty. The major sources of uncertainty of the result of measurement were identified and the combined uncertainty was calculated. Identification of the main uncertainty sources represents the basis for target operations for reducing the measurement uncertainty of this determination.
Keywords:Ammonium nitrogen  Wastewater analysis  Measurement uncertainty  Traceability
本文献已被 PubMed SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号