Estimation of measurement uncertainty for the determination of nonylphenol in water using solid-phase extraction and solid-phase microextraction procedures |
| |
Abstract: | We describe an estimation of measurement uncertainty calculated by the “bottom-up” approach for the determination of the oestrogenic compound nonylphenol in treated water samples by solid-phase extraction (SPE) and solid-phase microextraction (SPME) procedures and GC/MS detection. The results were compared and the different contributions to the uncertainty were evaluated. A study of the linear range was established and validation was performed for both methods using statistical analysis of several indicative parameters. In terms of validation data, precision (R.S.D. values <20%) and trueness (relative error <11%) were obtained for both methods under day-to-day conditions. The results of the estimation of measurement uncertainty obtained for both methods for concentrations higher than 1 μg/l have demonstrated that the time-consuming SPE method has a lower relative uncertainty (32%) than the SPME method (42.8%). The chromatographic uncertainty value was the main factor in the SPME method whereas the recovery factor (used to calculate the concentration) was the main contribution to uncertainty in the SPE method. |
| |
Keywords: | Uncertainty Validation data Water analysis Solid-phase extraction Solid-phase microextraction Nonylphenol |
本文献已被 ScienceDirect 等数据库收录! |